Pressure Systems, Inc. NetScanner™ System (9016, 9021, & 9022) User’s Manual
ii dd
Description ‘q’ read
index
www.PressureSystems.com
3 - 71
0B
00
01
Enable Automatic Shifting of Calibration Valve during
Calculate and Set Offsets (‘h’) command (default). (9016
only)
Disable Automatic Shifting of Calibration Valve in ‘h’.
User will manually control calibration value. (9016 only)
0C 00
01
Set Cal Valves to RUN or LEAK Position (default)
— choice made by ii=12. (9016 only)
Set Cal Valves to CAL/RE-ZERO or PURGE Position
— choice made by ii=12. (9016 only)
0B
see
chart
below
0D-0E -- Reserved for factory use
0F 00
01
Disable periodic Thermal Coefficient Update task.
Enable periodic Thermal Coefficient Update task
(default).
10 01-20
Set Number of A/D Samples to Average. (default = 8).
Valid values are 1, 2, 4, 8,10, 20 hex: (1, 2, 4, 8,16, 32
decimal)
05
11 -- Reserved for factory use
12 00
01
Set Cal Valves to RUN or CAL/RE-ZERO Position
(default) — choice made by ii=0C. (9016 only)
Set Cal Valves to PURGE or LEAK Position — choice
made by ii=0C. (9016 only)
0B
see
chart
below
13 00
01
Use Static IP Address Resolution (default)
Use Dynamic IP Address Resolution (RARP/BOOTP)
(Results in immediately becoming the module’s new
power-on default)
06
14 00
01
02
Disable Host Response/Stream Back-Off Delay (default).
Enable Host Response/Stream Back-Off Delay as low-
order byte of Ethernet Address(converted to decimal value
* 20 µsec.).
Enable Host Response/Stream Back-Off Delay specified
per eeee as decimal value (* 20 µsec.)