Pressure Systems 9022 Manuel d'utilisateur Page 128

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 194
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 127
Pressure Systems, Inc. NetScanner™ System (9016, 9021, & 9022) User’s Manual
www.PressureSystems.com
4 - 18
The example command structure will return the requested coefficient in ASCII floating point
format.
Reference
Coefficient
9021/9022
Command
for gain = 1
u01104
for gain = 20
u01105
for gain = 45
u01106
for gain = 90
u01107
for temperature
u01118
3
Commands to write the respective gain coefficient commands are summarized in the following
table. The example command structure will provide the new coefficient, shown as x.xxxx in ASCII
floating point format.
Reference
Coefficient
9021/9022
Command
for gain = 1
v01104 x.xxxx
for gain = 20
v01105 x.xxxx
for gain = 45
v01106 x.xxxx
for gain = 90
v01107 x.xxxx
for temperature
V0118 x.xxxx