Pressure Systems 9022 Manuel d'utilisateur Page 101

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 194
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 100
www.meas-
spec.com
Measurement Specialties Inc. NetScanner System (9016, 9021, & 9022) User’s Manual
1 - 68
Examples:
Sen
d TCP/IP command to module (i.e., via its connected socket): with
replacement values for the channel’s offset and gain correction terms loaded into
the module’s volatile memory (cc=00-01). Load these into channel # 8's
Transducer Coefficient array (aa=08).
v00800-01 0.000 1.000
Response returned is:
A
Send command to module (via its connected socket) to change its default EU
output from psi to kPa. This will be done by changing the EU Pressure
Conversion Scaler to 6.894757.
v01101 6.894757
Response returned is
Vue de la page 100
1 2 ... 96 97 98 99 100 101 102 103 104 105 106 ... 193 194

Commentaires sur ces manuels

Pas de commentaire