A special single Other Coefficients array is selected with array index aa=11 (hex).
All the valid coefficient indexes (for this array only) are listed in the following table:
cc
Other Coefficients Description
Datum
reserved - EU conversion offset term
EU Pressure Conversion scaler (default=1.0)
Reserved - EU conversion Non-Linearity term
Reserved-Reference Voltage value
Gain=1 Reference Coefficient (9021/9022 only)
Gain=20 Reference Coefficient (9021/9022 only)
Gain=45 Reference Coefficient (9021/9022 only)
07
Gain=90 Reference Coefficient (9021/9022 only)
FLOAT
18
Temperature Gain Reference Coefficient (9022
only)
FLOAT
Example:
• Send TCP/IP command to module (via its connected socket) requesting the most
recent calibration adjustment’s offset and gain terms (cc=00-01), and the adjacent
factory-determined transducer coefficients C0 through C4 (cc=02-06) for
transducer 1: Data requested in ASCII-hex format representing the internal binary
floating point format.
“u10100-06”
Response returned is:
“ 3B200A6E . . 00000000”
NOTE: The maximum response size is 300 characters. If the requested
range of coefficients requested exceeds this, the module will return an “N07"
error response.