Measurement Specialties Inc. NetScanner System (9016, 9021, & 9022) User’s Manual
3 - 15
RESET (Command ‘B’)
Purpose: Instructs the module to reset internal operating parameters, and to set all internal
control variables to their default “reset” state (see description below). The current
TCP/IP socket connection will remain open. Execution after a power off/on cycle
is optional (unnecessary).
Command
B
‘Bis the command letter.
Response
A
A’ is the acknowledge letter.
Description: The module returns to the following “reset” states if this command is executed:
Re-zero correction (offset) terms are set to the last values stored in
transducer memory.
Span correction (gain) terms are set to the last values stored in
transducer memory.
Calibration Valve is set to the RUN Position (9016 only).
Number of Samples for Data Averaging is set to last value stored in
non-volatile memory (factory default = 8).
Any autonomous host data delivery streams defined by c’
sub-commands are reset (undefined).
The Multi-Point Calibration function defined by C’
sub-commands is reset (undefined) if in progress.
Example:
Send TCP/IP command to a module (via open socket) to reset defaults:
B
Read following response:
A