www.meas-
spec.com
Measurement Specialties Inc. NetScanner System (9016, 9021, & 9022) User’s Manual
1 - 1
Chapter 1
General Information
1.1 Introduction
This Users Manual will:
Explain the electrical and pneumatic pressure connections for the NetScanner
System
Models 9016, 9021, and 9022 Intelligent Pressure Scanners.
Instruct you on how to program each module with computer software.
Instruct you on using the start-up software to manipulate and acquire data from each
module.
Model 9016 is a pneumatic Intelligent Pressure Scanner, with integral pressure transducers and a
pneumatic calibration manifold.
Models 9021 and 9022 are all-media Intelligent Pressure Scanners which may be fitted with up
to twelve (12) external all-media transducers (9400, 9401, 9402, or third party). Because of the
external nature of these transducers, and the variety of pneumatic or hydraulic media supported,
the 9021/9022 do not contain an integral calibration manifold.
Both models provide engineering unit pressure data with guaranteed system accuracy. This is
achieved by reading factory-determined pressure and temperature engineering-unit data
conversion coefficients from their transducers’ non-volatile memories at power-up. They also
allow additional adjustment coefficients to be “fine-tuned” with a multi-point calibration under
host control (e.g., possibly utilizing optional 903x Pressure Calibrator modules).
Models 9016, 9021 and 9022 provide 10-Base-T Ethernet communications for their Host Port
(with TCP/UDP/IP protocol).