Measurement Specialties Inc. NetScanner System (9016, 9021, & 9022) User’s Manual
4 - 2
manifold. (Models 9021 and 9022 have no valves.) Through software commands to each 9016, this
valve may be placed in one of four positions: RUN, CAL, PURGE, or LEAK-CHARGE. (See
the Set/Do Operating Options/Functions (‘w’) command (ii=0C & 12) in Chapter 3, and in
particular the Valve Position Chart at the end of command’s description.) When each module’s
internal calibration valve is placed in the CAL/RE-ZERO position (through software commands),
all DH200 transducer pressure inputs are pneumatically connected to the CAL input ports. All
DH200 reference inputs are pneumatically connected to the CAL REF input port. The CAL
input may be used to perform on-line zero adjustment of the 9016’s transducers. This capability
virtually eliminates sensor zero drift error and ensures the highest possible data accuracy. The
CAL input may also be used for DH200 span adjustment calibrations. Span calibration of
multi-range scanners may also utilize the CAL input ports if the highest applied pressure does not
exceed the proof pressure rating of any installed transducer. Zero and span adjustment functions
may also be performed on Models 9021 and 9022 channels, but any valving must be supplied
externally.
For reference when operating the 9016 calibration manifold, Figures 4.1 through 4.4 show
simplified pneumatic diagrams of the calibration manifold in its various positions.
NOTE: Periodic zero and span calibration should be the only calibration
required to maintain specified performance throughout the life of the scanner.